Commissioning; Heating Up; Temperature Setting - Vapormed Volcano MEDIC 2 Instructions For Use Manual

Hide thumbs Also See for Volcano MEDIC 2:
Table of Contents

Advertisement

5 . T H E VOLCA NO M E DIC 2 H O T A I R G E N E R AT O R

5.4. Commissioning

Do not leave the device unat-
tended while it is being opera-
ted. Put the Hot Air Generator
on a flat and stable surface.
Check whether the mains voltage cor-
responds to the voltage rating stated
on the device. The Power Cord and
the plug must be in perfect condition.
Only connect the Hot Air Generator to
a properly installed socket.
The device is safely disconnected
from the mains only when the plug is
removed from the socket.
When the power plug is plugged in, all
sequences of the LED display will il-
luminate briefly before displaying the
firmware number. The mains opera-
tion display which then appears on the
display shows that the Hot Air Genera-
tor is connected to the power supply.

5.5. Heating Up

To switch on the heater, press the
touch button labeled 'HEAT'. The set/
actual temperature will light up, signal-
ing that the heater is switched on. The
upper, orange display shows the ac-
tual temperature in the Filling Chamber
with the air pump switched on, and the
lower, white display shows the desired
set temperature. The heating process
is completed when the set value is
equal to the actual value. This is also
indicated acoustically when the pump
is switched on briefly. Heating up can
take up to two minutes depending
upon the chosen temperature. After
the heating process is completed, the
temperature is constantly maintained
at the set value.
For technical reasons the tem-
perature in the Filling Chamber
cannot be measured during
vaporization. During operation the
temperature is measured at the heat
exchanger. The display indicates the
reference values for the attached Fill-
ing Chamber with the air pump
switched on.
24
5 . T H E VOLCA NO M E DIC 2 H O T A I R G E N E R AT O R

5.6. Temperature Setting

Actual Value Display
(orange)
Mains Operation
Display
Touch Button
Temperature Minus
Touch Button
Heater (HEAT)
Pump Display
When you briefly tap on the touch but-
ton Temperature Plus or Minus, the set
values are increased or decreased in
the respective direction in 1-degree
increments.
If you keep the touch button Tempera-
ture Plus or Minus pressed, the set val-
ues start running continuously in the
corresponding direction. If the touch
button Temperature Plus and Minus is
activated simultaneously, the display
changes from degree Celsius to de-
gree Fahrenheit and vice versa.
If the actual value is higher
than the set value, you can
speed up the cooling down
process by switching on the air pump.
30 minutes after the last time a
button is pressed, the devi-ce
automatically switches off.
Set Temperature Display
Touch Button
Temperature Plus
Touch Button
Pump (AIR)
As long as the heater has a substan-
tial amount of residual warmth, the
temperature is displayed until it drops
below +40°C (+104°F).
Then the display turns off, and the on
the mains operation display indicates
that the Hot Air Generator is connect-
ed to the mains supply.
The device is safely discon-
nected from the mains only
when the plug is removed
from the socket.
25
(white)

Advertisement