Mount into the vessel
Mounting into the basin
Wall mounting
VEGABAR 66 • Foundation Fieldbus
4.6

Mounting steps with housing and thread

Fig. 10: Housing and thread
1
Housing
2
Seal
3
Thread
Mount VEGABAR 66 with housing and thread in the following way:
1. Weld the welded socket G1½ or 1½ NPT to the vessel top
2. Insert the transmitter with connection tube or suspension cable
into the opening
3. Turn the thread with seal into the socket and tighten with SW 46
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Mount VEGABAR 66 with housing and thread in the following way:
1. Fasten the mounting bracket at the suitable height on the basin
wall
Information:
We recommend articles from the line of VEGA accessories:
Mounting bracket of stainless steel, article no. 2.21615
Counter nut of PP, article no. 2.10371
2. Insert the transmitter with connection tube or suspension cable
into the opening of the mounting bracket and counter nut
3. Fasten counter nut with SW 46 on the thread
Disassembly is carried out in reverse order.
4.7

Mounting steps, external housing

1. Mark the holes according to the following drilling template
2. Depending on the mounting surface, fasten the wall mounting
plate with 4 screws
Seal with resistant material with thread 1½ NPT.
1)
4 Mounting
1)