Product temperature, suspension cable/measuring cell seal
Ʋ PE/FKM
Ʋ PE/EPDM
Ʋ PUR/FKM
Ʋ PUR/EPDM
Ʋ FEP/FKM
Ʋ FEP/FFKM
Ʋ FEP/EPDM
Product temperature, connection tube/seal meas. cell
Ʋ FKM
Ʋ FFKM
Ʋ EPDM
Product temperature, transmitter protection/seal meas. cell
Ʋ PVDF/FKM or EPDM
Ʋ PVDF/FFKM
Ʋ PE/FKM or EPDM
Calibration position
Influence of the installation position
Vibration resistance
With version with connection tube, vibration resistance 1 g. With lengths > 0.5 m (1.64 ft) the tube
must be supported in addition.
Shock resistance
Electromechanical data - version IP 66/IP 67
Cable entry/plug
17)
Ʋ Single chamber housing
Tested according to the guidelines of German Lloyd, GL directive 2.
15)
Tested according to EN 60068-2-27.
16)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.
17)
VEGABAR 66 • Foundation Fieldbus
-20 ... +60 °C (-4 ... +140 °F)
-20 ... +60 °C (-4 ... +140 °F)
-20 ... +80 °C (-4 ... +176 °F)
-20 ... +80 °C (-4 ... +176 °F)
-20 ... +100 °C (-4 ... +212 °F)
-10 ... +100 °C (+14 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-10 ... +100 °C (+14 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-20 ... +60 °C (-4 ... +140 °F)
-10 ... +60 °C (+14 ... +140 °F)
-20 ... +60 °C (-4 ... +140 °F)
upright, diaphragm points downward
< 0.2 mbar/20 Pa (0.003 psig)
mechanical vibrations with 4 g and 5 ... 100 Hz
Acceleration 100 g/6 ms
– 1 x cable gland M20 x 1.5 (cable: ø 5 ... 9 mm), 1 x
blind plug M20 x 1.5
or:
– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
– 1 x plug (depending on the version), 1 x blind stopper
M20 x 1.5
or:
– 2 x blind plug M20 x 1,5
10 Supplement