7.7

RadEye's battery status is lower than expected

Please check if the correct battery type is selected under "set-
tings" / "Batt. type" and change the type in case. Rechargeable
batteries supply a voltage of 1.2V each, while Alkaline batter-
ies provide 1.5V.
Rechargeable batteries offer also a more steep drop from ac-
ceptable battery voltage to "battery low" than Alkaline types.
7.8

History data shows the wrong time and date

See chapter 7-5.
7-6
DB-083 E
RadEye GF-Series
Le/Ff 20.07.2015