INSTALLATION AND OPERATION
4. Run the program by typing on the Controller "RUN <RETURN>".
2·40.
Interface Functions
The Generator implements a subset of interface functions defined by the I EEE Standard
488-1978. Table 2-9 summari�es the interface functions implemented. This section
describes the operation of the Generator in response to interface messages associated with
each interface function.
FUNCTION
SH1
AH1
TS
TEO
L3
LEO
SR1
RL 1
PPO
D C 1
DT1
co
E1
2-41 .
Address Mode
I n the address mode, the Generator may be operated from local (using the Front Panel
keys) or from remote (using the IEEE-488 Interface). The following paragraphs describe
the operation of the Generator in both states and transitions between the states.
The available IEEE-488 messages and their descriptions for the address mode of
operation are presented in Table 2-10.
2-42.
LOCAL OPERATION
The Generator powers up in the local mode. When
conditions arc present:
The front panel indicator REM. is not lit.
Device trigger (GET), device clear (DCL), and selected device clear (SOC) interface
messages are ignored.
All device dependent messages are ignored.
If the data output was requested while the Generator was in the remote mode, the
data output of a talker may be sent.
2-43.
GOING FROM LOCAL TO REMOTE
The Generator switches from local to remote when the "my listen address message"
(MLA) is received. and the Remote Enable (REN) signal is true.
2-38
Table 2-9. IEEE-488 Interface Functions List
D E S C R I PTION
Complete source handshake capabi l i ty
Complete acceptor handshake capabi l i t y
6as i c t a l ke r . Ta l k only, Ser i a l po l l � Unaddressed i f MLA
No ext ended t a l ke r capabi l i ty
Bas i c l i stenerr Listen only. Una ddressed i f MTA
No extended l i stener capabi l i ty
Complete serv i ce request c.apabi L i ty
Compl et e remote/ local capabi l i t y
No pa r a l lel po l l capa b i l i ty
t r i gger
Complete dev i ce c l ea r capabi l i t y
Complete device
capabi l i t y
No contro l l e r capabi l i ty
Open- c o l lector dri vers
m
local mode. the following
I