MAINTENANCE
PERFORMANCE TESTS
d. Adjust the Spectrum Analyzer to display the UUT signal for a convenient
e. Disconnect the Amplified rom UUT and terminate UUT OUTPUT with type-N
f.
g. Program the UUT to + 1 3 dBm.
h. Verify that the leakage is less than - 1 07dBm (I JJ.V), as indicated by the
1.
4A-10.
RF leakage is verified by measuring the induced signal in a two-turn loop with the
Measuring Receiver.
REQUIREMENTS
The output signal leakage must induce less than I JJ.V into a l-inch diameter two-turn
loop, I inch away from any surface of the generator as measured into a 50-ohm receiver.
TEST EQUI PMENT
Measuring Receiver
Two-Turn Loop
Type-N Termination
A screen room may be required depending on the RF environment.
REMARKS
This test is an alternative to the Output Leakage test.
Failing this test indicates a problem feed-through filter or improper mechanical
assembly, i.e., loose screws, and/ or damaged or misplaced gaskets.
The Measuring Receiver is used to measure the U UT leakage relative to a I
established at each frequency.
PROCEDURE
a. Connect the UUT 1 0 MHz OUT to the 10-MHz timebase input of the
b. Set the U UT REF !NT j EXT switch to INT.
c. Program the UUT to the Instrument Preset State, [RCL] [98).
d. Program the UUT to - 107 dBm.
4A-1 0
reference, using a vertical scale of 10 dB/ division, a resolution bandwidth of 3
kHz, and a span; division of
Termination.
Connect the two-turn loop to the Amplifier input.
Spectrum Analyz.er by moving the two-turn loop over the UUT surface at a
distance of I inch.
Repeat steps c through h at 550, 850, and 1050 M Hz.
ALTERNATE OUTPUT LEAKAGE TEST
Measuring Receiver.
Connect the Measuring Receiver sensor to the UUT RF OUTPUT.
5
kHz./ division.
I
I'
V reference