INSTALLATION AND OPERATION
Related Operations
Amplitude Fixed Range
Bright-Digit Edit Entry
Relative Function
Step Entry
2-23.
f
The
ollowing information describes how to use the Fixed-Range special function. This
special function fixes the current amplitude range (holds the currently selected step of the
Step Attenuator). This function allows monotonic and non transient level control ove
limited range around those levels where the Step Attenuator normally auto ranges. This
level control may be accomplished with the Bright-Digit Edit Entry only.
The level vernier in fixed range has at least 1 2 dB of range.
Command Syntax
Select Fixed Range -- Enable or Disable
Summary
Ena b l e F i xed Range
D i sa b l e F i xed Range
Example
Set the Generator for monotonic and non transient amplitude control (Bright-Digit Edit
only) over the range of the vernier level control below 0.25V.
Loc a l :
Remot e :
Notes
1 . The a
2.
Related Operations
Bright-Digit Edit Ent
Relative Functio
2-16
Amplitude Fixed Range
COMMAND
Loc a l ;
CSPCLJ C9J
C 1 J
Remot e ; " S P " "9"
" 1 "
Loc a l :
[SPCU
[9]
[Q]
''0''
Remote:
"SP" "9''
[AMPLJ
[ . ]
[2]
[5]
' ' A P . 2 5 V ; S P91 "
m
u
plit
de range is fixed only for Bright-Digit Edit operations. Other
methods of changing the amplitude cause the step attenuator to
necessary.
With amplitude fixed range d isabled, amplitude ranging occurs at 1 I 2V, 1 I 4 V,
1 18V, ...
12" V with AM off and l 14V, 1 1 8V, I / 1 6V, ... l 1 224V with AM on.