Establishing the daily operating reference density
levels for mammography application
(DT 2 Mammo) (optional)
This procedure enables you to establish the base line values for:
Low density
Mid density
High density
Procedure
1
Press the Key-operator key to enter the Key-operator mode.
2
Press the down key seven times, followed by the confirm key to select 'Quality
Control'.
5 Restore config.
6 Calibration
7 Service Actions
8 Quality Control
9 Installation
A confirmation screen appears:
108
WARNING:
The densitometer of the Drystar AXYS is calibrated at
installation. Authorized service personnel should recalibrate the
internal densitometer annually or after major service or repair.
2852 A EN 20070205