Performing the daily QC test for general
radiography applications (DT 2 B and DT 2 C)
Procedure
1
Turn on the Drystar AXYS and wait at least for 15 minutes. Refer to
on the Drystar AXYS"
2
Press the Key-operator key to enter the Key-operator mode.
3
Press the down key seven times, followed by the ok key to select 'QC'.
5 Restore config.
6 Calibration
7 Service Actions
8 Quality Control
9 Installation
A confirmation screen appears:
4
Do one of the following:
•
Press the Confirm key (YES) to start the QC procedure and proceed with step 5.
•
Press the Escape key (NO) to quit.
5
Define the tray for printing.
The 'Select tray' screen appears:
1.Upper tray
2.Lower tray
2852 A EN 20070205
WARNING:
This test must be performed every day before any clinical film
can be processed.
on page 48.