Karl Storz HAMOUT ENDOMAT Instruction Manual page 59

Table of Contents

Advertisement

51
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Recommended separation distances between portable and mobile RF communications equipment and the
®
®
The HAMOU
ENDOMAT
model 263310 20-1 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications
equipment.
150 kHz to 80 MHz
Rated maximum output
D
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re ection by structures, objects and
people.
Electromagnetic
Compatibility (EMC)
Information
Table 6
HAMOU
®
ENDOMAT
®
model 263310 20-1
Separation distance d [m]
according to frequency of transmitter
80 MHz to 800 MHz
D
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
(
)
800 MHz to 2.5 GHz
F
0.23
0.73
2.3
7.3
23

Hide quick links: