Guidance and manufacturer's declaration – electromagnetic immunity
®
HAMOU
ENDOMAT
The customer or user of the HAMOU
Immunity test
IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
Voltage dips, short
<5% U
interruptions and voltage
(>95% dip in U
variations on power supply
for 0.5 cycle
input lines
40% U
IEC 61000-4-11
(60% dip in U
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Electromagnetic
Compatibility (EMC)
Information
Table 2
®
model 263310 20-1 is intended for use in the electromagnetic environment specified below.
®
®
ENDOMAT
model 26 3310 20-1 should ensure that it is used in such an environment.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
*
<5% U
*
T
T
)
(>95% dip in U
T
for 0.5 cycle
40% U
T
T
)
(60% dip in U
T
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
T
for 25 cycles
<5% U
T
T
)
(>95% dip in U
T
for 5 sec
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
)
commercial or hospital environment.
T
If the user of the equipment or system requires
continued operation during power mains interruptions,
)
it is recommended that the equipment or system be
powered from an uninterruptible power supply.
)
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
(
)