30
The following table lists machine problem symptoms and possible areas to check for the problem causes. If these
suggested measures do not correct the machine malfunction, call your ACP Customer Support representative.
PROBLEM
Power Supply not plugged in to the
Unit will not power on
(under line power
unit of AC outlet
Power Supply not operational
use)
No program selected
Unit will not start
Cannot set carrier
Treatment in progress
frequency
Absent, inadequate, or improper
conductive medium interface
Lead wire(s): short or breakage
Patient feels surging
Non-conductive or poorly
or spiking sensation
during Estim.
conductive electrodes
Breakthrough effect of stimulation in
probe modes of operation
Failure of lead wire (s), electrode(s),
Patient cannot detect
or conductive medium interface
output during Estim
Failure of the Neuroprobe
SLD Infrared Pads
Red Diodes - some
Defective diodes
light up others are
dim or out
SLD Infrared Pads
Defective unit or pad
All of the diodes are
inoperative
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
TROUBLESHOOTING
CAUSE
Verify if Power Supply is connected as appropriate
Verify AC outlet is functional
Verify power plug used is appropriate and undamaged (see
figure 1 below)
Inspect Power Supply operation per procedure below
Select mode of operation: Infrared therapy, sensory/motor,
Stop treatment, adjust, and re-start
Replace with correct and adequate conductive medium
LEAD WIRE TESTER: Plug the distal (patient end) of the lead
wire into the test connectors while observing polarity (red /
black). Plug the device end of the lead wire into the receptacle
labeled LEAD WIRE TESTER. When the POWER is on, green
LED will turn on if the lead wire passes the test. LED will be
red if the lead wires are faulty and need to be replaced. (see
Figure 7. below)
Remove electrode(s) and replace if necessary
Adjust output up or down as necessary
Use Neuroprobe
LEAD WIRE TESTER: Plug the distal (patient end) of the lead
®
500 Pro
If Output Tester shows the unit to be functional, and the Lead
Contact ACP at (800) 350-1100 to replace the SLD Infrared
Pad
Try another pad to determine if the same condition exists. If
neither pad lights up contact ACP at (800) 350-1100 to replace
the Neuroprobe
contact ACP at (800) 350-1100 to replace the SLD Infrared
Pad
NEUROPROBE
REMEDY
noxious probe, nerve block
®
Output Tester to determine if unit has failed
or is operating incorrectly. Plug the Output Tester into the
Adapter. Turn on the Neuroprobe
If LED illuminates properly, test lead wire(s). (see figure 2
below) If LED is not illuminated, contact ACP Service Center
at (800) 350-1100.
wire into the test connectors while observing polarity (red /
black). Plug the device end of the lead wire into the receptacle
labeled LEAD WIRE TESTER.
green LED will turn on if the lead wire passes the test. LED
will be red if the lead wires are faulty and need to be replaced.
Wire tester show the lead wires to be functional, examine the
electrodes, or conductive medium interface for problems.
®
Unit. If one pad lights and the other does not
®
500 PRO USER MANUAL
®
. Increase output intensity.
When the POWER is on,