Accelerated Care Plus Neuroprobe 500 Pro User Manual page 21

Electrotherapy / infrared therapy system
Table of Contents

Advertisement

20
Remove the electrodes from their packaging. Apply the electrodes over the treatment sites points according to the
electrode placement techniques described in this manual. Various sizes of electrodes are available dependent upon
the size of the area to be treated. Follow the enclosed infection control procedures. Review the warnings and
application directions on the electrode packaging.
NOTE:
The use of conductive mediums (other than specifically approved pre-gelled or self adhering
electrodes) such as ultrasound gel or lotion, hand or body lotion, electrolyte spray mist, paper towels, non-
approved reusable or disposable pre-gelled or self-adhering electrodes, all increase the risk of irritation or
burns and are contraindicated for use with Neuroprobe
Inspect the full length of the lead wires for signs of frayed or cut wires and loose connections. Insert the stereo plug
completely into the device. Allow the lead wires to hang freely with no excessive strain on the stereo plug insulator.
Ensure that the electrodes are securely attached to the lead wires.
Periodically check the lead wires by using the LEAD WIRE TESTER on the front panel of the Neuroprobe
Insert the device-end of the lead wire into the connector labeled LEAD WIRE TESTER. Insert the patient-end of the lead
wires into the connectors within the white box and to the left of the device-end plug-in, matching the color coding. When
all wires are plugged in, the LED will turn green. If the LED is red, the cables are not inserted properly, or are faulty.
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
ACP Electrodes
®
Systems.
Lead Wires
Lead Wire Testing
®
NEUROPROBE
500 PRO USER MANUAL
®
500 Pro.