Accelerated Care Plus Neuroprobe 500 Pro User Manual page 32

Electrotherapy / infrared therapy system
Table of Contents

Advertisement

®
NEUROPROBE
500 PRO USER MANUAL
Figure 1
NOTE: When using the Neuroprobe
significantly illuminate the LED. Also, view the LED located at the end of the output tester directly. If viewed at an
angle, the LED may be too dim to notice that it is illuminated.
INSPECT POWER SUPPLY OPERATION
1.
Connect the power cord to the power supply and to a power outlet.
2.
Connect the power supply cable to the stimulator unit (Figure 4).
3.
Press the power supply switch ON (if applicable, Figure 5a).
NOTE: Newer power supply model does not have a ON/OFF switch (Figure 5b).
Figure 5a
4.
Verify the CHARGE light is on or blinking (Figure 6).
5.
Power on the stimulator and verify that you hear a click of the safety circuit relay energizing.
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
"HOSPITAL
GRADE"
marking on
the power
plug
LED
®
Output Tester, increase unit output intensity to 40 mA or higher in order to
Connect
power supply
cable to
stimulator
Press power
supply switch
ON
Output Tester
Figure 2
Figure 4
Figure 5b
31