The bar graph on the lower right of the display shows the relative outputs of the two channels with respect
to Vector operation. The numeric displays on the lower left show the output intensities into a 500 ohm load
in the output A/B modes. Each time a treatment is commenced, the output current level is reset to zero to
ensure that a patient cannot receive an initial shock due to a current output that was inadvertently left at a
high setting from a previous treatment. The PATIENT SAFETY switch, when pressed, also resets the
outputs to zero. Adjust the outputs to the appropriate level based on the protocol selected: sensory (pleasant
buzzing), motor (muscle twitch) or nerve block (numb gripping sensation).
Adjust intensity to desired level by pressing the up/down buttons for channels Output A and Output B (4).
During operation the parameters may be altered within the program at will by pressing the SELECT
PARAMETER button to select the parameter to be modified, then pressing the up/down buttons of the
MODIFY TIME/PARAMETER to modify the parameter as appropriate.
To stop the treatment push the START / STOP button (3) once.
NOTE: When zero is reached, a warning tone is emitted from the unit; the treatment current is then switched off
and T
O
REATMENT
unattended, a warning tone emits before the power automatically switches off. Note that, for all IFC or
INFRARED PAD protocols, the treatment time is preset but may be adjusted at will during the treatment.
Once the treatment is completed, remove the electrodes from the patient and disconnect the lead wires from
the electrodes.
Turn off the unit by pressing the POWER button (1/5).

Combination Therapy with IFC and SLD Infrared Pads

The SLD Infrared Pads may be used simultaneously with any of the Sensory Motor or Nerve Block pad therapy
protocols.
Press the POWER button (1/5) to turn-on the Neuroprobe
Plug in the SLD Infrared Pad into the output socket for the hand held treatment probe or SLD pad.
Apply the SLD Infrared Pad to the desired treatment areas.
Apply the stimulation electrodes to the treatment area.
Select the treatment modality desired and the appropriate treatment program by using the modality buttons
(2) to select Infrared Therapy in combination with Sensory / Motor, or Nerve Block programs.
Connect the electrodes to the lead wires and plug into the Neuroprobe
Push the START/STOP button (3) and start the treatments according to the instructions for IFC therapy
with pad electrodes.
Adjust intensity to desired level by pressing the up/down buttons for channels Output A and Output B (4).
To stop the treatment, push the START/STOP button (3) once.
NOTE: When zero is reached, a warning tone is emitted from the unit; the treatment current is then
switched off and TREATMENT OFF is displayed. Press the START/STOP button to shut off the warning
tone. If the unit is left unattended, a warning tone emits before the power automatically switches off. Note
that, for all IFC or INFRARED PAD protocols, the treatment time is preset but may be adjusted at will
during the treatment.
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
is displayed. Press the S
FF
NEUROPROBE
/ S
button to shut off the warning tone. If the unit is left
TART
TOP
®
500 Pro.
®
®
500 PRO USER MANUAL
500 Pro.