Summary of Contents for Accelerated Care Plus Neuroprobe 500 Pro

  • Page 2: User Manual

    PART NO. 72266 290NPRB Rev 4 This manual and all content therein is owned exclusively by Accelerated Care Plus ("ACP") and is protected by copyright. This manual or any portion thereof may not be photocopied, reproduced or translated to another language without the express prior written consent of ACP.
  • Page 3 ACP is internationally recognized for its contribution to research in the development of medical applications for therapeutic rehabilitation. The Company sponsors and conducts research at leading health care institutions and major universities throughout the world. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 4: Symbols On The Product

    Alternating current device. household waste. Follow state and/or local regulations. Observe the user manual Consult instructions for use Protect the product from ON / OFF push button humidity COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 5: Table Of Contents

    STANDARD LIMITED PRODUCT WARRANTY ................39 Warranty Coverage ............................39 Warranty Exclusion ............................39 Warranty Period ............................39 Warranty Validation ............................39 Return of Defective Equipment ........................40 COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 6: Electrotherapy/Infrared Indications & Contraindications

    Please notice that Accelerated Care Plus cannot provide medical advice. If you have specific medical questions, please contact your healthcare professional.
  • Page 7: Electrotherapy/Infrared Warnings & Precautions

    Please note that Accelerated Care Plus cannot provide medical advice. If you have specific medical questions, please contact your healthcare professional.
  • Page 8: Infrared Therapy Warnings

    If the patient continues to complain of pain, discontinue the treatment and shorten the treatment time on the next treatment session, or use an alternative type of therapy or electrode placement. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 9: Infrared Precautions

    Lay the SLD Infrared Pads flat whenever possible. Avoid excessive or unnecessary bending.  Do not apply SLD Infrared Pads to the body if wet. Dry the area thoroughly. The electronics in the pad may be short circuited if wet. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 10: The Neuroprobe ® 500 Pro

    500 Pro provides a nerve block mode through treatment pads with high output capable of sensory nerve block, or through the optional probe to produce local trigger point anesthesia. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 11: Controls And Functions

    H. PATIENT SAFETY switch. This is a remote duplicate of the START/STOP button. This is used by patients for treatment interruption. PATIENT GROUND connection. Patient ground pads for use in conjunction with the Point Stimulation Probe. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 12: Factory Settings

    The smaller the number, the faster the response time. Range is 0 to 99. Audio ON or OFF disables or enables the audio system which beeps at the end of the treatment. COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED CONFIDENTIAL AND PROPRIETARY...
  • Page 13: Power Supply Operation

    SLD INFRARED PADS plug into the same output socket as the probe.  To start the treatment, push the START/STOP button (3) once. The following screen will be displayed. INFRARED THERAPY COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 14: Ifc Therapy With Pad Electrodes

    (3). The treatment timer will begin, and the screen will provide basic treatment parameters, and will read as follows (i.e. if Pads Nerve Block was selected): IFC FF 5.0 KHz A: 00 mA B: 00 mA COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED CONFIDENTIAL AND PROPRIETARY...
  • Page 15: Combination Therapy With Ifc And Sld Infrared Pads

    If the unit is left unattended, a warning tone emits before the power automatically switches off. Note that, for all IFC or INFRARED PAD protocols, the treatment time is preset but may be adjusted at will during the treatment. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 16: Probe Mode Point Stimulation (Optional)

    Caution should be used with noxious, point stimulation. Patients may be intolerant or loss of sensitivity may mask stimulation levels that could result in skin breakdown or burns. COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED CONFIDENTIAL AND PROPRIETARY...
  • Page 17: Treatment Guidelines

    (higher percent duty cycle, higher average current intensity, and wider pulse widths) can significantly increase blood flow by altering the metabolic activity of muscles. PLOT OF IMPEDANCE AS A FUNCTION OF FREQUENCY USING EPIDUCTIVE SYSTEMS COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 18: Electrode Application Techniques

    MONOPOLAR PLACEMENT OF ELECTRODES USING HAND HELD PROBE COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED CONFIDENTIAL AND PROPRIETARY...
  • Page 19 This ensures that the current will intersect and thus provide an interferential pattern. The technique produces an interference field and is known as Interferential Current therapy of IFC. CERVICAL QUADRIPOLAR PLACEMENT OF ELECTRODES COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 20: Sld Infrared Pad Application Techniques

    Because of the increased current density available with pulsed or continuous medium frequency currents, a proportionally greater degree of conductive medium interface problems exists and should be monitored by the clinician. COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED CONFIDENTIAL AND PROPRIETARY...
  • Page 21 When all wires are plugged in, the LED will turn green. If the LED is red, the cables are not inserted properly, or are faulty. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 22: Infection Control Equipment And Principles Of Use

    M. tuberculosis as well as most bacteria and viruses. This will also facilitate removal of organic material contaminants from equipment.  Disposable /reusable electrodes are for individual patient use only and should not be used on multiple patients. COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED CONFIDENTIAL AND PROPRIETARY...
  • Page 23: Intermediate Level Disinfection

    Barriers should also be used on the equipment for treatment of non-intact skin or incontinence management. This technique will inactivate M. tuberculosis as well as most bacteria and viruses. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 24: Use Of Barriers - Intermediate Level Disinfection

    Factor based on the interface medium used. I.e. for use through a Tegaderm dressing, the treatment time would be increased from 15 minutes to 17 minutes to maintain the same dose. COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED CONFIDENTIAL AND PROPRIETARY...
  • Page 25 MODES OF OPERATION: Protocol reference sheet PROTOCOL APPLICATION Reference Sheet PROTOCOL / CLINICAL INDICATION STIMULATION PARAMETERS MECHANISM OF ACTION APPLICATION TECHNIQUE  45.4KHz, 62.5KHz, and 83.3KHz INFRARED THERAPY Target tissue - superficial and deep. Apply over or through the painful area or over the ...
  • Page 26: Neuroprobe ® 500 Pro Program Menu

    .0 Hz to 999 Hz 0 – 50% Burst rate scan: 0 – 20 Seconds Burst scan time: 10 Seconds Modulation: ON or OFF 0 – 999 Seconds Stim time: 60 Seconds COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 27 Burst scan time: 10 seconds Beat: Sweep: Rate Scan Time: Modulation: ON or OFF OFF, Fast 90º, Slow 90º, Vector: Fast 45º, Slow 45º 0 – 20% Auto Intensity: COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 28 ON or OFF Vector: 0 – 20% Auto Intensity: 0 – 99 Minutes Treatment time: 30 minutes Waveform: Burst Alternating Current 0 – 99 mA Output: 0 mA COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 29: Probe Noxious

    The black wires are of the same phase and the red wires in the opposite phase relative to each channel. All variable parameters may be altered during operation. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 30 ADJUSTABLE VALUES Carrier frequency: 5.0 KHz 0 – 99 Minutes Treatment time: 15 Minutes 0 – 20% Auto intensity: Waveform: Alternating Current 0 – 70 mA Output: 0 mA COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 31: Troubleshooting

    All of the diodes are the Neuroprobe Unit. If one pad lights and the other does not inoperative contact ACP at (800) 350-1100 to replace the SLD Infrared COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 32 Figure 5b Verify the CHARGE light is on or blinking (Figure 6). Power on the stimulator and verify that you hear a click of the safety circuit relay energizing. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 33: Service Center

    LED lights green when lead wire is functional, and lights red when lead wire needs to be replaced. Patient-end of lead wire, Black-coded Figure 7 Service Center COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 34 6:00am to 5:00pm Pacific Standard Time. After hours, please leave a message and a technician will return your call during the next scheduled workday. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 35: Specifications

    0-1.5 mA Peak current into 10K ohm load INFRARED THERAPY MODE: Waveform: Square wave 50% duty factor Pulse Rate: 45.4, 62.5and 83.3 KHz at 50% duty factor 1 min per frequency COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 36 Federal law restricts this device to sale by or on the order of a physician (or other health Caution: practitioner licensed by their state). NOTE: ACP reserves the right to change technical specifications and product availability without notice. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 37: Standard And Optional Accessories

    500 Pro Leatherette Soft Carrying * 67652 Case with ACP Logo ® 72266 Neuroprobe 500 Pro Operator Manual * This item is an optional accessory and is not included with the unit. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 38: Sld Infrared Pads

    Alligator Clip Assembly with Lead – Pair 22097 (1ea Black & 1ea Red) (for use with Single-use Electrodes) Alligator Clip Assembly, Pair 13539 (1ea Black & 1ea Red) (for use with Single-use Electrodes) COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 39: Infection Control Supplies

    1,200’ / roll ® Super Sani-Cloth Wipes, Single Use Packets 55536 (50 pkt/bx) ® 44425 Super Sani-Cloth Wipes, Tub (160/Tub) ® Sani-Cloth Wipes w/ Bleach, Tub 96849 (75 wipes/tub) COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 40: Standard Limited Product Warranty

    Buyer name or account number as it appears under the “Bill TO” on the ACP or recognized ACP Dealer invoice. Invoice Date and Number. Model number, description, and serial number of equipment. Detailed description of the problem. COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 41: Return Of Defective Equipment

    AUTHORIZATION TO DO SO. EQUIPMENT SENT IN WITHOUT AUTHORIZATION FROM ACP CUSTOMER SUPPORT WILL NOT BE ACCEPTED. Returned Materials Shipping Address: Accelerated Care Plus Attn: ACP Service Center 4999 Aircenter Circle, Suite 103 Reno, NV 89502 COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...