Accelerated Care Plus OMNISTIM FX2 PRO User Manual

Electrotherapy system

Advertisement

Advertisement

Table of Contents
loading

Summary of Contents for Accelerated Care Plus OMNISTIM FX2 PRO

  • Page 2: User Manual

    Part No. 290FX2C Rev 4 This manual and all content therein is owned exclusively by Accelerated Care Plus ("ACP") and is protected by copyright. This manual or any portion thereof may not be photocopied, reproduced or translated to another language without the express prior written consent of ACP. This manual may only be used by entities that have purchased the equipment or have implemented the ACP program and are covered by an executed lease agreement.
  • Page 3 The Company sponsors and conducts research at leading health care institutions and major universities throughout the world. This new medical frontier holds great promise and opportunity, which will result in substantial advancements in the health care industry and for ACP. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 4: Symbols On The Product

    Alternating current device. household waste. Follow state and/or local regulations. Observe the user manual Consult instructions for use Protect the product from ON / OFF push button humidity COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 5: Table Of Contents

    Service Center ..............................43 TECHNICAL SPECIFICATIONS .......................... 44 ® OMNISTIM PRO STANDARD AND OPTIONAL ACCESSORIES ............46 Electrodes ................................. 47 Infection Control Supplies ..........................47 STANDARD LIMITED PRODUCT WARRANTY ....................48 COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 6: Electrotherapy Indications & Contraindications

    Please notice that Accelerated Care Plus cannot provide medical advice. If you have specific medical questions, please contact your healthcare professional.
  • Page 7: Electrotherapy Warnings & Precautions

    Please note that Accelerated Care Plus cannot provide medical advice. If you have specific medical questions, please contact your healthcare professional.
  • Page 8: Precautions

    If there is a medical necessity to perform such treatments, these patients should be monitored diligently during application. The effect of electrical stimulation may be altered by the presence of these materials on the patient’s skin. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 9 Skin should be inspected after removal of electrodes for any signs of tearing or irritation.  Do not connect the stimulator to any electrical equipment for combination therapy except the Omnisound® family of ultrasounds. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 10: The Omnistim Fx Pro

    Pro MFAC and LVPC modes with fully adjustable ON and OFF Times and ON and OFF Ramps provides a wide variety of uses for muscle re-education and muscle spasm reduction programs for innervated muscle. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 11: Controls And Functions

    H. Connector for the device end of the lead wire when testing lead wire in the LEAD WIRE TESTER. AUX INPUT used to connect cable for remote hand control operation. PATIENT SAFETY input switch connector. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 12: Factory Settings

    The settings will reset to their initial default parameters if the batteries are removed or the unit is fully discharged. To restore the new settings follow the procedure in Appendix I. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 13: Battery Charger / Power Supply Operation

    The screen will display the following (in this example, TENS-LVPC MOTOR was selected from the NEURO RE-ED protocol functional group. NEURO RE-ED TENS-LVPC MOTOR SET OUTPUTS A-B A: 19 mA B: 19 mA 200 µS COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 14 Once the treatment is completed, remove the electrodes from the patient and disconnect the lead wires from the electrodes. Turn the unit off by pressing the POWER button (1/5). COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 15: Treatment Guidelines

    (i.e., higher percent duty cycle, higher average current intensity, and wider pulse widths), can significantly increase blood flow by altering the metabolic activity of muscles. PLOT OF IMPEDANCE AS A FUNCTION OF FREQUENCY USING EPIDUCTIVE SYSTEMS COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 16: Electrode Application Techniques

    This ensures that the current will intersect and thus provide and interferential pattern. CERVICAL QUADRIPOLAR PLACEMENT OF ELECTRODES COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 17 EXTENSOR PROFUNDUS DIGITI MINIMI • PATIENT IS PASSIVE DURING STIMULATION ELBOW FLEXION AND EXTENSION BICEPS BRACHII TRICEPS BRACHII CH A CH B • PATIENT IS PASSIVE DURING STIMULATION COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 18 ANTERIOR MIDLINE, MULTIFIDUS OVER THE MUSCLE CH B TRANSVERSE ABDOMINAL MUSCLE • PATIENT IS PASSIVE DURING STIMULATION • FOR UNILATERAL INVOLVEMENT, APPLY CH A TO THE MOST INVOLVED SIDE COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 19 PATIENT NEEDS TO BE SUPINE WITH HIPS EXTENDED KNEE FLEXION AND EXTENSION RECTUS FEMORIS VASTUS INTERMEDIALIS VASTUS LATERALIS BICEPS FEMORIS SEMIMEMBRANOSUS SEMITENDINOSUS VASTUS MEDIALIS CH B CH A (VMO) PATIENT IS PASSIVE DURING STIMULATION COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 20 PATIENT IS PASSIVE DURING STIMULATION IN A RELAXED TOE DOWN POSITION • TO ACCOMPLISH A TOE DOWN POSITION WHILE SEATED: PLACE A TOWEL ROLL UNDER HEEL, PLACE A PILLOW UNDER THIGH, RAISE SEAT HEIGHT COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 21 Insert the stereo plug completely. Allow the lead wires to hang freely with no excessive strain on the stereo plug insulator. ® Periodically check the lead wires by using the lead wire tester in the front of the Omnistim unit. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 22: Infection Control Equipment And Principles Of Use

    Barriers should also be used on the equipment for treatment of non-intact skin or incontinence management. This technique will inactivate M. tuberculosis as well as most bacteria and viruses. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 23: Use Of Barriers - Intermediate Level Disinfection

    10. Remove the plastic sleeve from the lead wire by sliding it toward the electrode. Remove the electrode and discard with the sleeve. ® 11. Intermediate disinfect the Omnistim Pro unit prior to the next treatment application. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 24: Modes Of Operaton: Protocol Reference Sheets

    MODES OF OPERATON: Protocol Reference Sheets Pain Control PROTOCOL / CLINICAL STIMULATION PARAMETERS MECHANISM OF ACTION APPLICATION TECHNIQUE OPERATIONAL SEQUENCE INDICATION  Press PAIN button until screen reads "IFC SENSORY". Sensory stimulation activates A-beta fibers Target tissue - superficial and deep. Bilateral, ...
  • Page 25: Neuro Re-Education

    Neuro Re-Education PROTOCOL / CLINICAL STIMULATION PARAMETERS MECHANISM OF ACTION APPLICATION TECHNIQUE OPERATIONAL SEQUENCE INDICATION This program is used to re-educate U.E. agonist and antagonist muscle pairs where fine motor  Press NEURO RE-ED button until screen reads "UE TRIPHASIC control and positioning is required.
  • Page 26: Muscle Re-Education

    Muscle Re-Education PROTOCOL / CLINICAL STIMULATION PARAMETERS MECHANISM OF ACTION APPLICATION TECHNIQUE OPERATIONAL SEQUENCE INDICATION  Press MUSCLE RE-ED button until screen reads "MFAC SURFACE Target tissue - superficial and deep muscle groups.  2500Hz - 75 BPS DEEP MUSCLE". Activation of muscle with electrical stimulation at a Bipolar set-up over agonist and antagonist muscles MFAC SURFACE DEEP...
  • Page 27: Omnistim Fx Pro Program Menu

    PHASE DURATION 0.0…250 (Hz) PULSE RATE 100 (Hz) PHASE DURATION MOD 20 (%) MODULATION RATE SCAN TIME 4 (s) 0…99 (min) TREATMENT TIME 15 (min) CHANNEL MODE A/B SIM. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 28 / trigger points. Quadripolar placement over area of local pain, or at involved spine segment. The intensity should be set to elicit a moderate muscle twitch. This protocol is comprised of three sub-programs, as follows: COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 29 BURST RATE SCAN 20 (%) RATE SCAN RATE SCAN TIME 15 (s) Slow 45, Fast 45, Slow 90, VECTOR Fast 90, OFF 0…99 (min) TREATMENT TIME 10 (min) COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 30: Neuro Re-Ed

    70 (μs) PHASE DURATION PULSE RATE 50 (Hz) 0…99 (min) TREATMENT TIME 15 (min) CYCLE TIME 2.5; 1.5; .75; .6; .5 (sec) .75 (sec) CHANNEL MODE Bi Phasic COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 31 PHASE DURATION 0.0…250 (Hz) PULSE RATE 4.0 (Hz) PHASE DURATION MOD 20 (%) MODULATION RATE SCAN TIME 4 (s) 0…99 (min) TREATMENT TIME 15 (min) CHANNEL MODE A/B SIM. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 32: Muscle Re-Ed

    A/B SIMULTANEOUS, A/B ALTERNATE, B DELAYED A/B SIM. When T-OFF = 0, T-ON = CONTINUOUS. B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 33: Functional Re-Ed

    Default (unit) Name Range (unit) Parameter Set Loop (Y/N) 40…100 (μs) 70 (μs) PHASE DURATION CYCLE TIME 1.40 (s) PULSE RATE 50 (Hz) 0…99 (min) TREATMENT TIME 15 (min) COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 34 Default (unit) Name Range (unit) Parameter Set Loop (Y/N) 40…100 (μs) 70 (μs) PHASE DURATION CYCLE TIME 1.0 (s) PULSE RATE 50 (Hz) 0…99 (min) TREATMENT TIME 15 (min) COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 35: Functional Re-Ed Assisted

    0.0…250 (Hz) BURST FREQUENCY 50 (Hz) BURST RATE SCAN 20 (%) RATE SCAN RATE SCAN TIME 4 (s) 0…99 (min) TREATMENT TIME 15 (min) CHANNEL MODE A/B SIM. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 36 PHASE DURATION 0.0…250 (Hz) PULSE RATE 50.0 (Hz) PHASE DURATION MOD 50 (%) MODULATION RATE SCAN TIME 4 (s) 0…99 (min) TREATMENT TIME 15 (min) CHANNEL MODE A/B SIM. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 37: Stimulation Therapy Modes

    The 2500 or 5000 Hz ® following technique is used in the Omnistim Pro Sport when in the full field mode of FULL FIELD WEDENSKY INHIBITION operation. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 38 When you wish to increase the amount of current density in the tissues to obtain a higher therapeutic dosage. When your electrode placement sites are less than optimal. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 39: Medium Frequency Alternating Current (Mfac)

    Gradually increase intensity to 15 to 30% OFF Ramp 2 Sec of maximum voluntary contractions. Continue the treatment for Pulse Rate 35Hz approximately 10 minutes. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 40: Low Voltage Pulsed Current (Lvpc)

    MFAC WAVEFORM Low Voltage Pulsed Current (LVPC) LOW VOLTAGE PULSED CURRENT WAVEFORM LVPC is generally used for surface and smaller muscle stimulation and pain control with TENS technique. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 41: Troubleshooting

    “HOSPITAL GRADE” marking on the power plug Output Tester Figure 1 Figure 2 COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 42 (Figure 4). Figure 4 Press the power supply switch ON (if applicable, Figure 5a). NOTE: Newer power supply model does not have a ON/OFF switch (Figure 5b). COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 43 LED is flickering between red to green, the lead wire should be replaced. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 44: Service Center

    6:00am to 5:00pm Pacific Standard Time. After hours, please leave a message and a technician will return your call during the next scheduled workday. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 45: Technical Specifications

    Scans amplitude of channel A relative to channel B which varies concurrently out of phase. PENS (PATTERNED ELECTRICAL NEUROMUSCULAR STIMULATION) MODE: Waveform: Asymmetric Biphasic Phase Duration: Variable from 40 µs to 100 µs Pulse Rate: Set at 50 Hz COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 46 CAUTION: Federal law restricts this device to sale by or on the order of a physician (or other health practitioner licensed by their State). ACP reserves the right to change technical specifications and product availability without notice. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 47: Omnistim Fx Pro Standard And Optional Accessories

    Leatherette Soft Carry Case Rechargeable Batteries (D size) 4400 mAh Heavy 65662 Duty Medical Grade Ni-Cd (Nickel Cadmium) (set of 4) ® 75542 Omnistim FX2 Pro User Manual COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 48: Electrodes

    Super Sani-Cloth Wipes, Tub (160 wipes/tub) ® Sani-Cloth Wipes w/ Bleach, Tub 96849 (75 wipes/tub) * This item is an optional accessory and is not included with the unit. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 49: Standard Limited Product Warranty

    Buyer name or account number as it appears under the “Bill TO” on the ACP or recognized ACP Dealer invoice. Invoice Date and Number. Model number, description, and serial number of equipment. Detailed description of the problem. COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...
  • Page 50 AUTHORIZATION. EQUIPMENT SENT IN WITHOUT AUTHORIZATION FROM ACP CUSTOMER SUPPORT WILL NOT BE ACCEPTED. Returned Materials Shipping Address: Accelerated Care Plus Attn: ACP Service Center 4999 Aircenter Circle, Suite 103 Reno, NV 89502 COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED...