NEUROPROBE
500 PRO USER MANUAL
Press the MODIFY TIME / PARAMETER button to set the desired treatment time if you wish to change
from the default settings of 15 minutes. Press the up or down button to increase or decrease the duration
from a minimum of one minute to a maximum of 99 minutes. The timer will count down to zero during
treatment.
NOTE: The SLD Infrared Pads always operate for the indicated treatment time for the selected protocol
with the same output regardless of the electrotherapy protocol chosen. The SLD Infrared Pads pulse at
sequential frequencies of 45.4, 62.5, 83.3, KHz at a 50% duty factor. Each frequency is active for 1 minute.
The frequency sequence is repeated until the timer terminates the treatment. This sequencing is not
discernable to the user.
To stop the treatment, push the START/STOP button (3) once.
When the treatment time is down to zero, the treatment current is switched off and INFRARED THERAPY is
displayed.
Turn off the unit by pressing the POWER button (1/5).
Press the POWER button (1/5) to turn on the Neuroprobe
the unit will be in the SELECT MODALITY/PROGRAM mode.
Select the treatment modality desired and the appropriate treatment program by using the modality buttons
(2) to select Sensory Motor, or Nerve Block pad programs.
Press the SELECT PARAMETERS button to select any settings to be modified.
TIME/PARAMETER button to adjust the desired treatment time, or other settings, if you wish to change from
the default settings. Time is preset to 15 or 30 minutes (dependent on the protocol selected). Press the up or
down buttons to increase or decrease the duration from a minimum of one minute to a maximum of 99 minutes.
Position the electrodes on the patient, connect the electrode lead wires and press the START/STOP button
(3). The treatment timer will begin, and the screen will provide basic treatment parameters, and will read
as follows (i.e. if Pads Nerve Block was selected):
COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED

IFC Therapy with Pad Electrodes

IFC FF
5.0 KHz
A: 00 mA
B: 00 mA
CONFIDENTIAL AND PROPRIETARY
®
500 Pro. The display screen will illuminate, and
T15
13