Select one of the Omnistim
appropriate lead wire from the electrostimulator and separate the two tip pins. Take the positive (red) tip pin and
insert it into the desired type and size of electrode, making certain that the electrode chosen is larger than the
transducer head size. This will allow the transducer to remain as the active electrode. Attach the self-adhesive
electrode to the patient's body. Take the negative (black) tip pin and insert it into the shrouded pin to banana
adapter (Item No. 2503036). Insert the adapter into the interconnect port of the Omnisound
®
Omnisound
transducer head will receive the designated output from the electrostimulator. Keep in mind that all
treatment parameters for electrostimulation (output intensity, frequency, treatment time, etc.) must be controlled
from the electrostimulator.
3000C Pro. Apply a liberal amount of conductive gel to the patient's body and begin combination
®
Omnisound
therapy through the Omnisound
COPYRIGHT © 2001 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED

Connection for Combination Therapy

®
family of electrostimulators for use with the Omnisound
However, all ultrasound treatment parameters will be controlled from the
®
transducer head.
®
OMNISOUND
3000C PRO USER MANUAL
®
3000C Pro. Choose the
®
3000C Pro. Now the