Sub-display parameters accuracy:
Ae = impedance (Z) accuracy
1
Definition: Q =
D
Rp = ESR (or Rs)× (1+
1. D value accuracy: De = ±Ae × (1+D)
2. ESR accuracy: Re= ±Z
ie., Z
= impedance calculated by
M
3. Phase angle θ accuracy: θe= ±(180/π) × Ae (deg)
Note: D:
dissipation factor
Q:
quality factor
ESR:
equivalent series resistance
Rp:
equivalent series parallel resistance
θ:
phase angle
)
× Ae (Ω)
M
or 2πf L