1)
According to the component type, press FUNC. key to select L, C or R measurement mode.
2)
Insert the standard component into the input terminal. Press SORTING key to enter into the sorting mode and the
"Sorting" symbol appears on LCD. If the LCD reading is OL or less than 200 counts, the SORTING key is not available.
3)
When sorting mode is active, press SETUP key to modify the range, reference value and the tolerance settings
sequentially.
"Range" symbol is flashing when setting the range. Press D/Q/ESR (←) key to shift the decimal point, unit to left and press
4)
SER/PAL(→) key to right. Press ENTER key to confirm and enter into the reference value setting mode automatically.
At this time, "Range" symbol disappear.
5)
When setting the reference value, press D/Q/ESR(←)key and SER/PAL(→)key to shift the bit to left and right
respectively. Press PCLINK (↑) key and REL% (↓) key to make the digit +1 or -1. The flashing bit is the current setting
bit. The reference value setting is available from 20 to 1999 counts. Press ENTER key to confirm and enter into the
tolerance setting mode automatically.
6)
When setting the tolerance, press D/Q/ESR(←)key and SER/PAL(→)key to select tolerance range: ±0.25% →±0.5%
→±1% →±2% →±5% →±10% →±20% →±80%-20%. The default tolerance is +1%. Press ENTER key to confirm.
7)
After setting the parameters, remove the standard component and insert the component to be measured. If the impedance
measured does not exceed tolerance range, the primary display will show "PASS" , otherwise show "FAIL". The current
measurement result will be shown on the secondary display.
8)
Press SORTING key again to exit the sorting mode.
Note: In AUTO LCR mode, the SORTING key is not available.