Summary of Contents for Aktakom AM?-3320

  • Page 1 AMМ-3320 Multi-function LCR USER`S MANUAL www.tmatlantic.com...
  • Page 2 АМM-3320 Thank you for purchasing the multi-function LCR from us. Please take a few minutes to browse through this user manual before you begin to operate the meter to ensure that you are fully familiarized with how best to operate the meter as accurately and safely as possible.
  • Page 3 АМM-3320 Introduction Features 19,999/1,999 counts dual LCD display  Basic accuracy: 0.3% with resolution of 0.01%  Analog bar display  Measurement frequency up to 100kHz  Mini-USB interface (USB model only)  AutoLCR smart check and measurement  Sorting function ...
  • Page 4 АМM-3320 Q:quality factor Rp:equivalent parallel resistance Frequency 100/120/1K/10k/100kHz Dual display + analog bar display Display 100/120 20mH~20kH 1kHz 2000uH~2000H 200uH~20H 100kHz 20uH~200mH 100/120 20nF~20mF 1kHz 2000pF~2mF 200pF~200uF Measurement 100kHz 200pF~20uF range 100/120 200Ω~200MΩ 1kHz 20Ω~200MΩ 20Ω~20MΩ 100kHz 20Ω~2MΩ 200~200M 0.001~1999 θ...
  • Page 5 АМM-3320 Calibration Open/Short function Interface Mini-USB Measurement Approx.1.2 times/second speed Measurement 4-terminal terminal Basic accuracy 0.3% Power AAA*6 battery or external power supply Auto power off 5 min(with batteries) temperature 0℃~40℃ Operating environment humidity ≤90%RH (40℃, no condensation) Storage -25℃~50℃ temperature Impedance accuracy Ae The below-listed accuracies are guaranteed by the meter with normal use under the operating temperature of...
  • Page 6 АМM-3320 10 – 100k – 1M – 20Μ− 0.1- 1Ω 1 – 10Ω Remark Freq. 100kΩ 1MΩ 20ΜΩ 200MΩ 1.0%+5d 0.5%+3d 0.3%+2d 0.5%+3d 1.0%+5d 2.0%+5d 100/ 1.0%+5d 0.5%+3d 0.3%+2d 0.5%+3d 1.0%+5d 2.0%+5d D < 0.1 120Hz 1kHz 1.0%+5d 0.5%+3d 0.3%+2d 0.5%+3d 1.0%+5d 5.0%+5d...
  • Page 7 АМM-3320 Sub-display parameters accuracy: Ae = impedance (Z) accuracy Definition: Q = Rp = ESR (or Rs)× (1+ 1. D value accuracy: De = ±Ae × (1+D) 2. ESR accuracy: Re= ±Z × Ae (Ω) ie., Z = impedance calculated by or 2πf L 3.
  • Page 8 АМM-3320 Explanation on Front Panel Figure 1 The front panel is shown as figure 1. 1. Mini USB interface Connect with PC, easily for data transmission and management. 2. LCD Used for displaying the measuring results and various symbols. Used for putting on or off the operating power for the meter. 4.
  • Page 9 АМM-3320 8. HOLD key Used to maintain the measurement data unchanging, by pressing the key again it will resume the measurement. 9. D/Q/ESR key In L/C measurement mode, press the key to select parameters of D/Q/θ/ESR. 10. SETUP key When sorting mode is active, press SETUP key to modify the reference value, range and the tolerance settings sequentially. 11.
  • Page 10: Enter Key

    АМM-3320 By pressing this key for once, the backlight of the LCD screen will be opened and after 60 seconds the meter will automatically turn off the backlight. It is also possible to turn off the backlight by pressing this key before the 60 seconds. 15.
  • Page 11 АМM-3320 Understanding Display Screen Figure 3 LCD screen is shown as in Figure 3, with its every symbol’s meaning shown as in the Table 1:...
  • Page 12: Sub-Display

    АМM-3320 Meaning Meaning Sub-display Auto power off indication Data hold Analog bar indication Auto mode indication The Meter is in the data transmission mode. Auto LCR mode indication battery indication (with battery supply) Range indication Frequency indication Tolerance range Relative Measurement mode. Main-display Sorting mode indication Unit for main parameters...
  • Page 13: Function Descriptions

    АМM-3320 Function Descriptions 1. Parameters setting Press FUNC. key to select the following parameters sequentially: AUTO LCR, L-Q, C-D, R, DCR. Parameter Meaning AUTO LCR Auto LCR smart mode Inductance measurement,the parameter on sub-display is quality factor Q. Capacitance measurement, the parameter on sub-display is dissipation factor D.
  • Page 14 АМM-3320 If θ > 11°, the Auto-L mode is selected. The parameter on sub-display is Q. If θ < -11°, the Auto-C mode is selected. The parameter on sub-display is D. If the C < 5pF, the parameter on sub-display is parallel resistance Rp. Note: In order to avoid damaging the instrument, the capacitance requires discharge before measuring.
  • Page 15 АМM-3320 6. Open/Short calibration 1)Press CAL key larger than 2 seconds to start the open/short calibration procedure. 2) In open calibration mode, the secondary display will show “Open”. There are two ways for open state input: a When using square terminals, the square terminals and L terminals hang in the air (shown in figure b When using L terminals,insert the black and red testing lines with alligator clip into the "L...
  • Page 16 АМM-3320 FAIL symbol will show on the primary display. Press CAL key again to save the calibration data and enter into the short calibration mode. 4) In short calibration mode, the secondary display will show “Srt”. There are two ways for short state input: a When using square terminals, insert the short socket to the square terminals and make L terminals hang in the air (shown in figure 6);...
  • Page 17 АМM-3320 FAIL symbol will show on the primary display. Press CAL key again to save the calibration data. Note: 1. To get the better accuracy, the open/short calibration should be done before measurement. 2. The purpose of open/short calibration is to reduce the parasitic effect of the test fixture. 3.
  • Page 18 АМM-3320 According to the component type, press FUNC. key to select L, C or R measurement mode. Insert the standard component into the input terminal. Press SORTING key to enter into the sorting mode and the "Sorting" symbol appears on LCD. If the LCD reading is OL or less than 200 counts, the SORTING key is not available. When sorting mode is active, press SETUP key to modify the range, reference value and the tolerance settings sequentially.
  • Page 19 АМM-3320 9. PC-LINK mode Press PCLINK key and USB symbol appears on LCD. Connect the instrument to PC through USB interface, and the measured data can be recorded, analyzed, processed and printed by PC. Press PCLINK key again to cancel the data transmission.
  • Page 20 АМM-3320 Figure 8 Figure 9 2. Capacitance measurement 1) Turn on the power. 2) If exists voltage in the capacity, connect the two ends of the capacitor for a short time to discharge. 3) There are two ways for capacitance input: a.
  • Page 21 АМM-3320 are not available. 5)Press FUNC. key twice to select Auto-C mode. The primary LCD display will show the capacitance value. The secondary quality factor Q, LCD display will show the dissipation factor D. The equivalent resistance ESR/Rp or phase angle θ can also be shown by pressing the D/Q/ESR key.
  • Page 22 АМM-3320 3. Resistance measurement 1) Turn on the power. 2)There are two ways for resistance input: a. Insert the measured resistance into the input terminal directly (shown in figure 12); b. Connect the alligator clips to the ends of the measured resistance (shown in figure 13). 3)The default test mode is Auto LCR mode, the resistance value will show on primary display and the phase angle θ...
  • Page 23 АМM-3320 Figure 12 Figure 13 To use Adapter (Under external power adapter, the automatic power off function is not available.) Connecting the power adapter: 1,Connect the AC power cord to the AC-DC converter. 2,Plug the AC power cord into an electrical outlet (100V-240V). 3,Plug the DC power plug of the converter into DC power socket of the meter.
  • Page 24 АМM-3320 DC power DC power AC power plug socket cord AC-DC converter. AC/DC adapter information: Input:100V-240VAC, 50-60Hz 1.8A Output: DC 12V 2A MAX Polarity: WARNING: 1. Please use the original AC power adapter, using other AC power adapter may damage your instrument. 2.
  • Page 25: Maintenance

    АМM-3320 8. Please make the AC power adapter avoid a strong bump. 9. It is normal when the AC power adapter make some noise in use. Maintenance Cleaning Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents. Calibration Calibrate your instrument once a year to ensure that it performs according to its specifications.
  • Page 26 АМM-3320 If there are some changes in accessories, please refer to the real product as standard. Note: To get the better accuracy, when using L terminals measured with the alligator clip-testing lines, Please insert the open socket to the square terminals (shown in figure 14). Figure 14...