2.7.2 Installing the low-volume vials

Warning:
To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
To install low-volume vials:
1.
If a standard reservoir bottle is fitted, remove it.
2.
Screw the low-volume adaptors into the manifold and finger-tighten them.
Figure 2–2:
Low-volume adaptor
Low-volume vial
Solvent delivery tube
3.
Screw the low-volume vials into the adaptors.
4.
For each low-volume vial, ensure that the ends of the solvent delivery tubes are positioned
so that they are close to, but do not touch, the bottom of the vial (see
delivery tube

2.7.3 Adjusting the solvent delivery tube positions

For correct operation of the IntelliStart Fluidics system, you must adjust each solvent delivery
tube so that its end is close to, but does not touch, the bottom of the reservoir bottle or low
volume vial.
Warning:
To avoid laceration injuries caused by the shattering of fragile, low-
volume glass vials, take care when screwing them in, and never use force.
positions).
June 9, 2016, 715004990 Rev. A
Page 42