Equivalent Sampling (Equ Smpl) - Iwatsu DS-8814 Operation Manual

Table of Contents

Advertisement

52
Equivalent
Using equivalent sampling for a repetition signal can speed up effective
sampling
sampling rate than the maximum single-shot sampling rate.
(EQU SMPL)
Principle
Every time waveform data is captured, the time tn (n=1, 2, 3,...) after
trigger generation until data is sampled is measured and data is
overwritten taking account of relative time relationship.
This instrument implements 10 G equivalent sampling by measuring
the time with 100 ps resolution.
Example of measurement by normal sampling and Equ sampling for
approx.100 MHz sine wave
Norm sampling
Sampling in 5 ns period
Mainly used to
measure single-shot signal.
Join setting in Display menu
This Norm sampling example is
shown by Join ON.
Data pieces captured in 5ns
period are linked with a line.
Trigger
5 ns period
Displays other than Equ
5 ns period
Equivalent sampling
100 ps equivalent sampling
Can displays detailed waveform.
Restriction on
equivalent sampling
With long memory, equivalent
sampling is not performed
even if Equ Smpl is set to ON.
3rd waveform data captured
1st waveform data captured
2nd waveform data captured
Displays Equ