Operating Vertical Axis; Offset, Volts/Div, Zoom - Iwatsu DS-8814 Operation Manual

Table of Contents

Advertisement

14

Operating vertical axis

MENU:
SELECT: Sets a channel, through which MENU, OFFSET, or VOLTS/
Trace
Selects display (On) or non-display (Off) of CH1/CH2/CH3/CH4 waveform.
OFFSET
Sets the position of the waveform in the vertical direction.
Pressing OFFSET moves the waveform in
1-division steps in the direction it was last
moved. The offset variable range is as follows:
2 mV to 50 mV/DIV:
100 mV to 500 mV/DIV:
1 V to 10 V/DIV:
VOLTS/DIV
Selects the deflection factor in 1-2-5 sequence.
This function varies depending on the offset setting in the CH menu.
Division
Enlarges or reduces the waveform based on the GND reference mark
(GND position).
GND position
Screen center
GND position
Volts
Enlarges or reduces the waveform based on the screen center.
Set the waveform to the screen center with the OFFSET knob. Useful
for observing the exact part of the waveform you are interested in.
Screen center
Zoom
Pressing the VOLTS/DIV knob allows the vertical gain sensitivity to be
alternately switched between 1-2-5 sequence and Zoom. (The same
function is available using the CH menu.)
If "Zoom" is selected, the setting varies continuously from ´1.00 to ´2.50
(maximum 0.8 mV/div). (Expansion by software)
The waveform is enlarged based on the screen center regardless of the
offset setting.
Displays the vertical menu.
DIV can be operated.
Used to observe waveform at other than
the center of the screen, for example,
to compare and measure two waveforms.
Used to observe waveform at the center of the screen,
for example, to measure a specific part of waveform or
waveform carrying DC bias.
±1 V
±10 V
±100 V
Turn clockwise (moves up)
Turn counterclockwise
(moves down)