Vertical Axis; Horizontal Axis - Iwatsu DS-8814 Operation Manual

Table of Contents

Advertisement

Vertical axis

Probe
Selects an attenuation ratio of the probe used by the probe setting in the
CH1 to CH4 menu from 1:1, 10:1, 100:1 and 1000:1.
setting
Selecting Auto automatically recognizes the probe with a probe sense
function (1/10, 1/100) and sets an attenuation ratio.
When using the probe, set Auto. The deflection factor, trigger level,
etc. are converted and displayed depending on the attenuation ratio.
The converted value is reflected in cursor measurement and auto
measurement, etc.

Horizontal axis

The setting range for a trigger delay in a waveform acquisition (RUN)
DELAY
state varies depending on the sweep time and memory length.
The waveform acquisition stop (STOP) state is −500 s to +750 s
regardless of sweep time and memory length.
In the RUN state, the time from the trigger
point to the center of the waveform screen
is set.
The STOP state is used to move an acquired waveform in horizontal direction.
See page 16 for Post trigger and Pre trigger.
Trigger delay setting range (RUN state)
Memory length
Short
Long
Sweep time
5 ns to 1 ms/div
2 ms to 50 s/div
5 ns to 20 ms/div
50 ms to 50 s/div
Trigger
Setting range
−5 div to +10 ms
−5 div to +500 s
−5 div to +200 ms
−5 div to +500 s
Screen
Delay
45
− 5 div

Advertisement

Table of Contents
loading

Table of Contents