Taking A Reading - Elcometer 204 User Manual

Steel ultrasonic material thickness gauge
Hide thumbs Also See for 204:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
5 SETTING THE ZERO POINT (continued)
To set the zero point:
1
Plug the transducer into the gauge
ensuring that the connector is fully
engaged.
The wearface of the transducer should
be clean and free of any debris.
2
Press the On/Off button to switch the
gauge on.
3
Press Menu/Calibration/Zero Set and
apply couplant to the zero disk.
4
When prompted, press the
transducer on to the zero disk,
making sure it is flat against the
surface.
The display will show a thickness value
which is constantly updating. The
stability of the reading is indicated on the
stability bar to the right of the display. A
valid reading has a stability of 5 or more.
5
Remove the transducer from the zero
disk. The last reading is held on
screen. If not representative, repeat
Step 4.
Excessive use of couplant can result in a
distorted reading when the transducer is
removed from the surface.
6
Press 'Zero' to set the zero point.

6 TAKING A READING

6.1 BEFORE YOU START
1
Press the On/Off button to switch the gauge on.
2
Connect a transducer to the gauge.
All dual element transducers which can be connected directly to the base
of the gauge - see Section 10.1 'Transducers' on page en-8 - are
'intelligent' transducers and will be identified automatically by the gauge.
If using other Elcometer 'non-intelligent' dual element transducers or other
manufacturers' transducers, a transducer adaptor is required - see
Section 10.4 'Transducer Adaptor' on page en-11.
3
Set the zero point of the transducer - see Section 5 on page en-5.
4
Prepare the test surface - see Appendix 1 on page en-14.
www.elcometer.com
R
en-6

Advertisement

Table of Contents
loading

Table of Contents