Power-On Resets; Lcd Backlight; Lcd Contrast; Remaining Device Lifetime - Siemens SITRANS LR560 Operating Instructions Manual

Radar
Hide thumbs Also See for SITRANS LR560:
Table of Contents

Advertisement

4.4. Power-on Resets

Read only. The number of power cycles that have occurred since manufacture.
To view via SIMATIC PDM, open the menu Device – Wear.

4.5. LCD Backlight

Time the backlight remains on.
Options

4.6. LCD Contrast

The factory setting is for optimum visibility at room temperature and in average light
conditions. Extremes of temperature will lessen the contrast.
Values
Adjust the value to improve visibility in different temperatures and light conditions.
Change the value in small steps to ensure you can continue to read the display.
To access LCD Contrast via SIMATIC PDM:
Navigate to Level Meter > Setup > Local Display.

4.7. Remaining Device Lifetime

Notes:
• Default settings in the parameter tables are indicated with an asterisk (*) unless
explicitly stated.
• Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and
set up Service/Calibration schedules, based on operating hours instead of a
calendar-based schedule. See also Remaining Sensor Lifetime (4.8.), Service
Schedule (4.9.), and Calibration Schedule (4.10.).
• Performing a reset to Factory Defaults will reset all the Maintenance Schedule
parameters to their factory defaults.
• The device operates in years. To view Remaining Device Lifetime parameters in
hours or days (via SIMATIC PDM only) see Lifetime Expected (4.7.1.).
The device tracks itself based on operating hours and monitors its predicted lifetime.
You can modify the expected device lifetime, set up schedules for maintenance
reminders, and acknowledge them.
The maintenance warnings and alarms are communicated via the Status byte. This
information can be integrated into any Asset Management system. For optimal use,
we recommend that you use SIMATIC PCS7 Asset Management Software in
conjunction with SIMATIC PDM.
Page 88
SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS
Range: 0 (backlight off) to 128 seconds (backlight always on).
Default: 128 seconds
Range: 0 to 20
Default: 8
Contrast setting will depend on ambient
temperature.
7ML19985LT01

Advertisement

Table of Contents
loading

Table of Contents