2.5.9.2.
Note: Fail-safe Mode (2.5.9.) must be set to Substitute Value before Value
(2.5.9.2.) can be defined.
User-defined default for the Output Value, if sensor or sensor electronic
fault is detected. Units are defined in Unit (2.5.8.2.).
Values

2.6. AIFB2

See AIFB1 (2.5.): the parameters for AIFB2 are identical.

2.7. Measured Values

Read only. Allows you to view measured values for diagnostic purposes
In SIMATIC PDM, open the menu View – Process Variables.
2.7.1. Main Output
The value for Level/Volume. (PV– Primary Value).
2.7.2. Output, no linearization
The value for Level (SV1 – Secondary Value 1).
2.7.3. Output, no level offset
The value for Distance (SV2 – Secondary Value 2).

2.8. Filtering

2.8.1. Narrow Echo Filter
Filters out echoes of a specific width.
Values
To remove a false echo from the Echo Profile, take its width in mm and multiply
it by 0.013. Enter the result.
For example, to filter out a spike with 500 mm width, enter 6 or 7 (the closest
integer product of 500 x 0.013).
When a value is keyed in, the nearest acceptable value is entered.
2.8.2. Reform Echo
Smooths jagged peaks in the echo profile. Reforms fragmented echoes into one
echo.
Values
7ML19985LT01
SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS
Value
Range: -999999 to 999999
Default: 0
Range: 0 to 255. Default: 0.
0 = OFF
greater = wider
Range: 0 to 255 samples.
Default: 0 (reset to 10 after a Quick Start has been completed)
0 = OFF
greater = wider
Recommended: 5 to 20 samples. Wider is not recommended.