To assign a diagnosis to a particular event via the handheld programmer:
a) Go to Enable (3.5.1.) and select Yes to enable Condensed Mode.
b) Go to Event Index (3.6.1.) and enter the event index number corresponding
to a particular event.
c) Go to Event Diagnosis (3.6.3.) and choose a Diagnosis option from the table
above.
To assign a status to a particular event via SIMATIC PDM:
a) Go to Level Meter > Maintenance & Diagnostics > Condensed Status >
Enable, and select Yes to enable Condensed Status Mode.
b) Go to Level Meter > Maintenance & Diagnostics > Allocation Level.
c) Select the Diagnosis line for the selected Event, then choose a Diagnosis
option from the associated pull-down menu.
3.7. Peak Values
To view via SIMATIC PDM:
Open the menu View – Device Diagnostics, select Device Status, and click on the tab
Device Status. For more details see
3.7.1. Min. Measured Value
The minimum recorded Sensor value, reported in units defined in Unit (2.3.1.).
3.7.2. Max. Measured Value
The maximum recorded Sensor value, reported in units defined in Unit (2.3.1.).
3.7.3. Minimum Output Value - AIFB1
The minimum recorded Output Value from the Analog Input Function Block 1.
3.7.4. Maximum Output Value - AIFB1
The maximum recorded Output Value from the Analog Input Function Block 1.
3.7.5. Minimum Output Value - AIFB2
The minimum recorded Output Value from the Analog Input Function Block 2.
3.7.6. Maximum Output Value - AIFB2
The maximum recorded Output Value from the Analog Input Function Block 2.
4. Service
Note: Default settings in the parameter tables are indicated with an asterisk (*)
unless explicitly stated.
4.1. Demo Mode
Reduces the time between measurements and the accuracy for demonstration
purposes.
Page 86
SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS