Siemens SITRANS LR560 Operating Instructions Manual page 124

Radar
Hide thumbs Also See for SITRANS LR560:
Table of Contents

Advertisement

CLEF Range (2.4.5.4.)
Determines the level below which the CLEF algorithm will be used. Above this level the
Center algorithm is used when Hybrid is selected in Position Detect (2.4.5.2.).
CLEF Range is referenced from Far Range.
CLEF Range
Echo Threshold (2.4.5.3.)
Confidence (2.4.7.1.) describes the quality of an echo. Higher values represent higher
quality. Echo Threshold defines the minimum confidence value required for an echo to be
accepted as valid and evaluated.
Echo Lock (2.4.6.1.)
If the echo selected by Algorithm is within the Echo Lock window, the window is centered
about the echo, which is used to derive the measurement. In radar applications, two
measurement verification options are used:
Lock Off
SITRANS LR560 responds immediately to a new selected echo (within the restrictions set
by the Maximum Fill / Empty Rate), but measurement reliability is affected.
Material Agitator
A new measurement outside the Echo Lock Window must meet the sampling criteria
before the window will move to include it.
The other available options, Maximum Verification and Total Lock are not recommended
for radar.
Auto False Echo Suppression (2.4.8.1.)
Notes:
• For detailed instructions on using this feature via PDM see
Suppression
• For detailed instructions on using this feature via the handheld programmer see
Auto False Echo Suppression (2.4.8.1.) on page 75.
Auto False Echo Suppression is designed to learn a specific environment (for example, a
particular vessel with known obstructions), and in conjunction with Auto False Echo
Suppression Range to remove false echoes appearing in front of the material echo.
Page 118
SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS
Low Calibration
Point
on page 48.
sensor
reference
point
Far
Range
Auto False Echo
7ML19985LT01