Maintenance; General Maintenance Information; Cleaning; Periodic Inspection - ConMed 1000 SES Service Manual

Smoke evacuation system
Table of Contents

Advertisement

3.1 General Maintenance Information

This section contains information for ordinary
maintenance of the CONMED 1000SES
Smoke Evacuation System. While the unit has
been designed and manufactured to high industry
standards, it is recommended that periodic inspec-
tion and performance testing be performed by a
qualified Biomedical Technician to ensure contin-
ued safe and effective operation.

3.2 Cleaning

Before cleaning the CONMED 1000SES
unplug the unit from the power outlet. The exte-
rior of the unit may be cleaned by wiping it with
a cloth that has been dampened (not dripping)
with a disinfectant or a mild soap or detergent
solution such as Windex® or Formula 409®.
Windex® is a registered trademark of the Drackett
Products Company. Formula 409® is a registered
trademark of the Clorox Corporation. Wipe dry
with a clean cloth. Do not steam sterilize.

Maintenance

Section 3.0

3.3 Periodic Inspection

The CONMED 1000SES
inspected at least every year. This inspection
TM
should include checks for:
•Damage to the power cord.
•Damage to the power plug.
•Tightness of the power plug.
•Proper mating, cleanliness and absence of
damage to the filter inlet.
•Obvious external or internal damage to the unit.
TM
,
See the following page for Troubleshooting.
should be visually
TM
3-1