Technical Data - Siemens MAG 1100 F Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for MAG 1100 F:
Table of Contents

Advertisement

SITRANS F M
2.

Technical data

2.1 Sensor MAG 1100 and MAG 1100 HT (High temperature)
Type
Measuring principle
Excitation frequency
(Main supply: 50 Hz / 60 Hz)
Process connection
Nominal size
• MAG 1100 (Ceramic)
• MAG 1100 (PFA)
Mating flanges
Rated operating conditions
Ambient conditions
1)
Ambient temperature
• Sensor
• Sensor ATEX
• Compact transmitter
MAG 5000/6000
Temperature of medium
• MAG 1100 (Ceramic)
• MAG 1100 ATEX (Ceramic)
• MAG 1100 (PFA)
Temperature shock
• MAG 1100 (Ceramic)
- Duration
1 min, followed
by 10 min rest
• MAG 1100 (PFA)
Operating pressure
• MAG 1100 (Ceramic)
• MAG 1100 (PFA)
Mechanical load
2. Technical data
MAG 1100
DN 2 ... 65 (
/
" ... 2½"): 12.5 Hz / 15 Hz
1
12
DN 80, 100 (3", 4"): 6.25 Hz / 7.5 Hz
DN 2 ... DN 100 (
/
" ... 4")
1
12
DN 10 ... DN 100 (
/
" ... 4")
3
8
EN 1092-1 (DIN 2501), ANSI B 16.5 class 150
and 300 or equivalent
Option:
DN 2 ... 10 (
/
" ...
/
"):
1
3
12
8
G½" / NPT ½" pipe connection adapters
-40 ... +100 °C (-40 ... +212 °F)
-20 ... +60 °C (-4 ... +140 °F)
-20 ... +50 °C (-4 ... +122 °F)
-20 ... +150 °C (-4 ... +302 °F)
-20 ... +150 °C (-4 ... +302 °F)
-30 ... +130 °C (-20 ... +266 °F)
Suitable for steam sterilization at 150 °C (302 °F)
• DN 2, 3 (
/
",
/
") No limitations
1
1
12
8
• DN 6, 10, 15, 25: Max. ∆T
", ½", 1": Max. ∆T
(¼",
/
3
8
• DN 40, 50: Max. ∆T
70 °C/min
(1½", 2", 2½": Max. ∆T
70 K/min)
• DN 80, 100: Max. ∆T
60 °C/min
(3", 4": Max. ∆T
60 K/min)
Max. ±100 °C (210 °F) momentarily
• DN 2 ... 65: 40 bar (
/
" ... 2½": 580 psi)
1
12
• DN 80: 37.5 bar (3": 540 psi)
• DN 100: 30 bar (4": 435 psi)
-6
Vacuum: 1 x 10
bar
(1.5 x 10
abs
20 bar (290 psi)
Vacuum: 0.02 bar
(0.3 psi
abs
DN 80 ... DN 100: CO
pressure max. 7 bar (101.5 psi)
2
• 18 ... 1000 Hz random in x, y, z, directions for
2 hours according to EN 60068-2-36
• Sensor: 3.17 grms
• Sensor with compact MAG 5000/ 6000 mounted
transmitter: 3.17 grms
MAG 1100 HT (High temperature)
Electromagnetic induction
DN 15 ... 50 (½" ... 2"): 12.5 Hz / 15 Hz
DN 80, 100 (3", 4"): 6.25 Hz / 7.5 Hz
DN 15 ... DN 100 (½
EN 1092-1 (DIN 2501), ANSI B 16.5 class 150
and 300 or equivalent
-40 ... +100 °C (-40 ... +212 °F)
-20 ... +50 °C (-4 ... +122 °F)
-20 ... +200 °C (-4 ... +392 °F)
-20 ... +180 °C (-4 ... +356 °F)
• DN 15, 25: Max. ∆T
80 °C/min
(½", 1": Max. ∆T
80 K/min)
• DN 40, 50, 65: Max. ∆T
(1½", 2": Max. ∆T
• DN 80, 100: Max. ∆T
(3", 4": Max. ∆T
• DN 15 ... 50: 40 bar (½" ... 2": 580 psi)
• DN 80: 37.5 bar (3": 540 psi)
• DN 100: 30 bar (4": 435 psi)
-5
psi
)
Vacuum: 1 x 10
abs
)
abs
• 18 ... 1000 Hz random in x, y, z, directions for
2 hours according to EN 60068-2-36
• Sensor: 3.17 grms
MAG 1100 and MAG 1100 Ex
"
... 4")
80 °C/min
80 K/min)
70 °C/min
70 K/min)
60 °C/min
60 K/min)
-6
-5
bar
(1.5 x 10
psi
)
abs
abs
9