SITRANS F M
(continued)
8.6.4
MAG 5000 CT and
MAG 6000 CT settings
Internal totalizers
Output
80
8. Commissioning
mm
[inches]
DN 600
24
DN 700
28
DN 750
30
DN 800
32
DN 900
36
DN 1000
40
42
DN 1100
44
DN 1200
48
DN 1400
54
DN1500
60
DN1600
66
DN1800
72
DN 2000
78
Setting primary operating parameters such as Q
blocked during normal operation. See menu setup.
These settings are made in connection with commissioning or calibration by mounting a hardware
key on the connection plate of the transmitter. When the key is mounted, there is access to all menu
items. When the key is removed, the primary settings are blocked in accordance with the
requirements in the authorisation.
Depending on the type of approval it is possible to reset the internal totalizers. The type of approval
is selected in the reset menu, with the hardware key mounted. It is possible to choose between:
Hot/cold water
Other liquids
Resetting of totalizers by electrical input is not possible.
Hot/cold water
Totalizer 1 is allocated to forward flow (cannot be reset).
Totalizer 2 is allocated to reverse flow (cannot be reset).
Other liquids
Both totalizer 1 and totalizer 2 are allocated to measure the net flow, i.e. any reverse flow will make
the totalizers count backwards.
Totalizer 1 cannot be reset.
Totalizer 2 can be reset if the flow velocity in the meter pipe is <0.25 m/s. When the totalizer is
reset, the pulse output register will also be reset.
When choosing hot water, changing the output settings is not allowed and the output settings
menus will not be shown in display.
When choosing cold water or other liquids, all output settings can be changed.
Volume/pulse or batch quantity
MAG 5100 W
(Order no.
7ME6520)
min.
max.
3
236 ml
6177 m
3
321 ml
8407 m
3
368 ml
9651 m
3
419 ml
10981 m
3
530 ml
13897 m
3
654 ml
17157 m
3
654 ml
17157 m
3
792 ml
20760 m
3
942 ml
24706 m
-
-
-
-
-
-
-
-
-
-
, low flow cut-off, units, approvals, etc. is
max.
MAG 1100, MAG 1100 F,
5100W
(Order no. 7ME6580)
MAG 3100, 3100 P
min.
max.
3
236 ml
6177 m
3
321 ml
8407 m
3
368 ml
9651 m
3
419 ml
10981 m
3
530 ml
13897 m
3
654 ml
17157 m
3
654 ml
17157 m
3
792 ml
20760 m
3
942 ml
24706 m
3
1.28 l
33628 m
3
1.15 l
38604 m
3
1.68 l
43923 m
3
2.12 l
55590 m
3
2.62 l
68629 m