5.3 Adjusting the display range
(function group BASE)
The function group BASE enables you to make the
basic adjustment of the display range. The display on
the screen must be adjusted for the material to be test-
ed (function MTLVEL) and for the probe used (function
P-DELAY).
– If required, go to the first operating level.
– Select the function group BASE.
5-6
H Note:
In order to accurately adjust the material velocity and
the probe delay, please read the section Calibrating the
USM 35X , chapter 5.7, beforehand.

RANGE (Display range)

You can adjust the range for your measurement in
RANGE.
• Coarse adjustment:
from 0.5 mm ... 1400 (9999) mm in even steps
• Fine adjustment:
up to 9.99 mm in steps of 0.01 mm
up to 99.9 mm in steps of 0.1 mm
up to 999 mm in steps of 1 mm
H Note:
The adjustment range for the display range depends on
the frequency range setting (function FREQU in func-
tion group RECV).
Issue 01, 04/2005