Interfaces and Peripherals
Function
Code Range/Default
PROBE-#**
PB
0 = variable / alphanumerical input
1 = B1-S
2 = B2-S
3 = B4-s
4 = MB2-S
5 = MB4-S
6 = MB5-S
7 = MWB45-2
8 = MWB60-2
9 = MWB70-2
10 = MWB45-4
11 = MWB60-4
12 = MWB70-4
RANGE
DW
0.5 - 4 MHz / 0.2 - 1 MHz:
0.5 - 9999 mm / 250
0.02 - 390" / 10
2 - 20 / 0.8 - 8 MHz:
0.5 - 1420 mm / 250
0.02 - 56" / 10
R
RATING
BD
-1101 - 1101 db
RECALL
RD
0 = off
1 = on
8-22
Resolution
13 = SWB45-2
14 = SWB60-2
15 = SWB70-2
16 = SWB45-5
17 = SWB60-5
18 = SWB70-5
19 = WB45-1
20 = WB60-1
21 = WB70-1
22 = WB45-2
23 = WB60-2
24 = WB70-2
0.01
0.001
0.01
0.001
0.1
1
Issue 01, 04/2005
Function
Code Range/Default
RECTIFY
RF
0 = full-w
1= pos hw
2 = neg hw
3 = rf
REFECHO
RC
0 = no ref.
1 = ref. stored
REFECHO**
RE
0 = BW
1 = SDH
2 = FDH
REFMODE
RO
0 = off
1 = on
REFRNCE
R
BB
0 - 1101 dB
REFSIZE**
RS
0.5 - 10 mm / 3
REJECT
RJ
0 - 80 % / 0
SCALE
SE
0 = measval
1 = snd-pth.
2 = div.
Remote control
Resolution
1
1
0.1
0.01
1