Evaluation according to the DGS method (only USM 35X S)

Selecting the DGS mode

– If required, go to the third operating level.
– Select the function group CFG2.
– Switch the function EVA-MOD over to the setting
DGS.
– Go to the second operating level.
– Select the function group DGS.
Krautkramer USM 35X
Double assignment of functions:
The following functions are double assigned. Toggle
between the two functions by repeatedly pressing the
corresponding key
DGSMEN>
Calling the DGS menu
T-CORR>
Setting the transfer
correction

Default settings for the DGS measurement

In the next step, the DGS menu is called enabling to
select the corresponding probe and to set the other
DGS parameters:
– Select the function DGSMEN.
– Use the right-hand rotary knob to call the DGS menu.
Issue 01, 04/2005
Operation
.
DGSMEN>
Switching on/off the
DGS evaluation mode
OFFSET>
Activating the multi-curve
display mode
5-73