Adjusting the display range (function group BASE)
Frequency range
Adjustment range (c = 5920 m/s)
0.2 to 1 MHz
0.5 to 9999 mm/0.02" to 390"
0.5 to 4 MHz
0.5 to 9999 mm/0.02" to 390"
0.8 to 8 MHz
0.5 to 1420 mm/0.02" to 50"
2 to 20 MHz
0.5 to 1420 mm/0.02" to 50"
– Select the function RANGE.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the required value by means of the right-hand
rotary knob.

MTLVEL (Sound velocity)

Use MTLVEL to set the sound velocity within the test
object. You can use sound velocities between 1000 and
15000 m/s.
• Coarse adjustment, in steps as follows (m/s):
15000
9000
14000
8000
13000
7000
12000
6320
11000
6000
10000
5920
Krautkramer USM 35X
5000
2000
4000
1600
3250
1450
3130
1000
3000
2730
Issue 01, 04/2005
• Fine adjustment
1000 ... 15000 in steps of 1 m/s
A
Attention:
Always ensure that the function MTLVEL is correctly
set. The USM 35X calculates all range and distance
indications on the basis of the value adjusted here.
– Select the function MTLVEL.
– If required, toggle between coarse and fine adjust-
ment.
– Adjust the required value by means of the right-hand
rotary knob.

D-DELAY (Display starting point)

Here you can choose whether to display the adjusted
range (for example 250 mm) starting from the surface
of the test object, or in a section of the test object
starting at a later point. This allows you to shift the
complete screen display and consequently also the
display zero.