3.5 B-SCAN Display Range

The default BSCAN display range is 0-20 inch. We highly recommend adjusting the range according to the target, because setting an
appropriate range will display a clearer scan graph.To adjust the range, complete the steps below.
Note: If the BSCAN value exceeds the preset range, that point will be graphed as equal to the range limit.
1.
Press
to access the Configuration display (see Figure 9 on page 21).
2.
Press
or
to select either BSCAN MINIMUM or BSCAN MAXIMUM.
3.
When the desired parameter is highlighted, press
4.
Press
or
to change the selected position in alarm thickness and then press
position. To switch the alarm setting to OFF from any other value, simultaneously press
5.
Press
after the alarm thickness has been set to the desired value.
DM5E Operating Manual
to activate it.
Chapter 3. Setting Up the DM5E
or
to adjust the value of the selected
or
.
31