General configuration
H Note:
Double assignment of the function DATE/TIME (icon >).
Toggle between the two functions by repeatedly press-
ing the corresponding key
A
Attention:
Always make sure that you are using correctly set
time and date values. Otherwise test results might be
corrupted. Be aware that the USM 35X displays the
year as a two digit number!
– Select the function TIME.
– Use the left-hand rotary knob to highlight the value
that you want to change, e.g. the hour.
– Use the right-hand rotary knob to change the high-
lighted value.
– Select the function DATE.
– Use the left-hand rotary knob to highlight the value
that you want to change, e.g. the day.
– Use the right-hand rotary knob to change the high-
lighted value.
Krautkramer USM 35X
.
Issue 01, 04/2005

ANAMOD

You can output results of measurements at the analog
output for external further processing. Use the function
ANAMOD to configure the analog output in case there
is no echo in the evaluation gate and the analog voltage
has been selected for the sound path at the output.
You have the following setting options:
• lo volt
The analog output supplies 0 volt.
• hi volt
The analog output supplies 5 volts.
– Select the function ANAMOD.
– Use the right-hand rotary knob to choose the required
value.
Operation
5-57