Initial start-up
The charging time is 10 hours with a simultaneous ul-
trasonic inspection. If the instrument is not being used
for ultrasonic inspections, the charging time is 8 hours.
This charging time applies to ambient temperatures
from 25 to 30 °C. Please take into consideration that
the batteries are not charged to their full capacity at
higher temperatures.
The LED display on the plug-in power supply unit indi-
cates the status of the charging process.
green LED
off
off
off
flashing dark/bright
on
off
off
3-6
yellow LED
flashing
flashing dark/bright
on
flashing bright/dark
off
off
off
Issue 01, 04/2005
External charging
Lithium-ion batteries can be charged by means of an
external battery charger. We recommend the battery
charger with the order number 35 297. To charge single
NiCad or NiMH cells, you need the external desktop
battery charger with the order number 101 729.
red LED
off
flashing bright/dark
off
off
off
flashing bright/dark
on
Power supply
Status
no battery detected
charging at low power
quick charging phase 1
quick charging phase 2
battery charged
temperature error, auto-reversible
csharging error, permanent
Krautkramer USM 35X