If You Cannot Restore Raw Files; If You Cannot Telnet To The Command Parser Socket - HP 1670A Series User Manual

Logic analyzers
Table of Contents

Advertisement

Troubleshooting
If you cannot restore raw files
If
you cannot restore raw files
If the setup. raw and data. raw files seem to be ignored when you copy
them to the logic analyzer:
D Verify the LAN connection between your computer and the logic
analyzer. Refer
to
"If you cannot connect to the logic analyzer" earlier in
this section.
D Change the file time-out. To change the time-out:
1 Go to the logic analyzer System External 1/0 menu.
2 Select the LAN Settings field.
3 Set the file time out.
A value of about 1.5 seconds should be sufficient. Set the time out
to
the minimum value that will allow network operations
to
work
reliably.
If
you cannot telnet to the command parser socket
If you get a "connection refused" message:
D Verify that you included the telnet port number (5025) in the command.
If you get a "connection timed out" or "no response from host" message:
D Verify the LAN connection between your computer and the logic
analyzer. Refer to "If you cannot connect to the logic analyzer" earlier in
this section.
If you get a "connection refused" or "no response from host" message:
D If the logic analyzer was just turned on, make sure that you wait 15
seconds after the Analyzer Configuration menu is displayed before
attempting the connection.
D Make sure that another user is not already accessing the file system as
the control user or connect.ed to the command parser socket.
7-24

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660cs series1660c series

Table of Contents