Electrical Installation (With Integral Electronics) - Emerson Rosemount Oxymitter 4000 Instruction Manual

Oxygen
Hide thumbs Also See for Rosemount Oxymitter 4000:
Table of Contents

Advertisement

Oxymitter 4000
Figure 3-8. Installation with Drip Loop and Insulation Removal
ELECTRICAL
INSTALLATION (WITH
INTEGRAL
ELECTRONICS)
3-10
For Oxymitter 4000 with Integral Electronics
All wiring must conform to local and national codes.
Disconnect and lock out power before connecting the power supply.
Install all protective covers and safety ground leads after installation. Failure to install covers
and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation
of this equipment, connection to the main electrical power supply must be made through a
circuit breaker (min 10 A) which will disconnect all current-carrying conductors during a fault
situation. This circuit breaker should also include a mechanically operated isolating switch.
If not, then another external means of disconnecting the supply from the equipment should
be located close by. Circuit breakers or switches must comply with a recognized standard
such as IEC 947.
Instruction Manual
IM-106-340, Rev. 4.3
July 2017