Semi-Automatic Calibration; Manual Calibration With Membrane Keypad - Emerson Rosemount Oxymitter 4000 Instruction Manual

Oxygen
Hide thumbs Also See for Rosemount Oxymitter 4000:
Table of Contents

Advertisement

Instruction Manual
IM-106-340, Rev. 4.3
July 2017
Semi-Automatic
Calibration
Manual Calibration with
Membrane Keypad
3. If using an IMPS 4000, enter a time interval via the IMPS 4000 keypad
that will initiate an automatic calibration at a scheduled time interval (in
hours). To set the CalIntvX parameter of the CHANGE PRESETS
display mode, refer to the IMPS 4000 Intelligent Multiprobe Test Gas
Sequencer Instruction Manual for more information.
Once an automatic calibration is initiated, by any of the methods previously
described, the Oxymitter 4000's CALIBRATION RECOMMENDED alarm
signals an IMPS 4000 or SPS 4001B to initiate a calibration. The sequencer
sends an "in cal" signal to the control room so that any automatic control loops
can be placed in manual. Then, the sequencer begins to sequence the
calibration gases.
Semi-automatic calibrations only require operator initiation. However, the
calibration gases must be permanently piped to the Oxymitter 4000, an SPS
4001B or IMPS 4000 must be installed to sequence the gases, and the logic
I/O must be set to mode 8 or 9 via HART/AMS to allow the sequencer and the
Oxymitter 4000 to communicate.
Depending on your system setup, a semi-automatic calibration can be
initiated by the following methods:
1. Oxymitter 4000 with membrane keypad. Press the CAL key on the
Oxymitter 4000 keypad.
2. Oxymitter 4000 with LOI. Select "Start Calib" from the CALIBRATION
menu.
3. IMPS 4000. Use the IMPS 4000 keypad to change the InitCalX
parameter of the CHANGE PRESETS display mode from 0000 to 0001.
Refer to the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer
Instruction Manual for more information.
4. HART. Use the HART Communicator to access the O
menu and perform the O
O
Cal Method" in Section 7: HART/AMS for the complete calibration
2
procedure.
5. AMS. Refer to AMS documentation for more information.
6. Remote Contact. Initiate a calibration from a remote location via the
remote contact input connection provided by an IMPS 4000 or SPS
4001B. Refer to the documentation available for the control system in
use for more information.
Once a semi-automatic calibration is initiated by any of the methods
previously described, the Oxymitter 4000's CALIBRATION RECOMMENDED
alarm signals an IMPS 4000 or SPS 4001B to initiate a calibration. The
sequencer sends an "in cal" signal to the control room so that any automatic
control loops can be placed in manual. Then, the sequencer begins to
sequence the calibration gases.
Manual calibrations must be performed at the Oxymitter 4000 site and will
require operator intervention throughout the process. Manual calibration
instructions, in condensed form, can also be found on the inside of the right
electronics housing cover. See Figure 9-2.
Use the following procedure to perform a manual calibration:
1. Place control loop in manual.
Oxymitter 4000
CAL method. Refer to "HART Communicator
2
CALIBRATE
2
9-3