Table 8-1. Diagnostic/Unit
Alarm Fault Definitions -
Membrane Keypad Only
LED
Flashes
HEATER T/C
1
2
3
4
HEATER
1
2
3
4
5
O
CELL
1
2
3
4
CALIBRATION
1
2
3
**
*Critical alarm conditions will render the O
selectable limit of 3.5 mA or 21.6 mA (position 3 of SW2). Factory default value is 3.5 mA. Alarms which are not self-clearing
(Self-Clearing = NO) will require a reset. Perform the Reset Procedure in Section 2: Configuration of Oxymitter 4000 with Membrane Keypad to
continue operation.
**The CALIBRATION RECOMMENDED alarm flashes the Calibration Recommended LED on the operator's keypad.
Table 8-2. Diagnostic/Unit
Alarm Fault Definitions - LOI
Message
O2 T/C OPEN
O2 T/C SHORTED
O2 T/C REVERSED
ADC ERROR
O2 HEATER OPEN
VERY HI O2 TEMP
BOARD TEMP HI
O2 TEMP LOW
O2 TEMP HI
O2 CELL OPEN
O2 CELL BAD
EEPROM CORRUPT
CALIB FAILED
LINE FREQ ERROR
8-6
Status
Open
Shorted
Reversed
A/D Comm Error
Open
High High Temp
High Case Temp
Low Temp
High Temp
High mV
Bad
EEprom Corrupt
Invalid Slope
Invalid Constant
Last Calibration Failed
Calibration Recommended
measurement as unusable, and any of these events will cause the 4-20 mA signal to go to a user-
2
Status
Heater T/C Open
Heater T/CShorted
Heater T/C Polarity Reversed
A/D Comm Error
O2 Heater Open
Very High Process Temperature
Electronics Overheated
Low Process Temperature
High Process Temperature
O2 Cell Open
O2 Cell Failed
EEprom Failed
Last Calibration Failed
Incorrect Input Line Frequency Detected on Power Up
4-20 mA Line
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
Track O
2
3.5 mA (factory default)*
Track O
2
Track O
2
Track O
2
Track O
2
Fault Number
1
2
3
4
5
6
7
8
9
10
11, 13, 14
12
15
Instruction Manual
IM-106-340, Rev. 4.3
July 2017
Fault
Recoverable
1
No
2
No
3
No
4
No
5
No
6
No
7
Yes
8
Yes
9
Yes
10
Yes
11
Yes
12
No
13
Yes
14
Yes
15
Yes
Yes
Self Clearing
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
No
Yes
No