Unit (Adjustment via <F1> - <F3>):
- Calculation of gas concentrations in:
- ppm = volume concentration (parts per million)
3
- mg/m
=
- mg/kWh = mass concentration per performance unit
Undiluted (Adjustment via <F4>):
- Conversion of gas concentration on
entered reference oxygen:
- Formula for conversion:
E
ref
O
reference (press <OK> to access input mode):
2
- Input of reference O
Air pressure (press von <OK> to access input mode):
- Input of barometrical air pressure for the dew point calculation
Set clock (press <OK> to access input mode):
- Correction of internal clock
Type of fuel (press <OK> to access selection mode):
- Modification of adjusted fuel type
Internal (press <OK> to open menu):
- Further instrument adjustments:
Keyboard beep (<F1> for Yes / <F4> for No):
- Acoustical signal by pressure on key
Baud rate (adjustment via cursor keys):
- Adjustment of transfer speed by data transfer via RS 232
(1200 -19200 Baud)
Online data (adjustment via cursor keys):
- Adjustment if online data should be sent via RS232 or IR
Page 14
mass concentration per volume unit
21 – O
= E
*
meas
21 – O
2ref
CO
undiluted
ppm mg/m3 mg/kwh
2ref
2meas
Keyboard beep
Baud rate
Online data
ECOM-P
Eta(C)
9999
ppm