13
Handling the camera

13.3 Adjusting the infrared camera focus

13.3.1 Figure

13.3.2 Procedure

Follow this procedure:
1. Do one of the following:
• For far focus, rotate the focus ring clockwise (with the touch-screen LCD facing to-
ward you).
• For near focus, rotate the focus ring counter-clockwise (with the touch-screen
LCD facing toward you).
Note Do not touch the lens surface when you adjust the infrared camera focus man-
ually. If this happens, clean the lens according to the instructions in 26.2 Infrared lens,
page 132.
Note It is very important to adjust the focus correctly. Incorrect focus adjustment affects
how the image modes Thermal MSX, Thermal, and Picture-in-picture work. It also affects
the temperature measurement.