Perform Setup
 To delete the stored data of the white balance calibration, tap DELETE
WHITE BALANCE DATA (2).
 After deleting the stored data, run the white balance again, as
described.
 If problems arise during the white balance calibration, contact Image
Access technical support, see section Technical Support starting at page
6.
 To return to the previous submenu, tap BACK (1).
 To return to the "Start screen", tap EXIT (3).
After a successful white balance, proceed as follows:
 Remove the test target.
 Put the transport guides back into their position.
 Store the test target in a place which is protected from daylight.
 Ensure that the test target is not damaged, bent or soiled.