1.3 Connections

1.2.4.1 Input overload
overload limits
1.2.4.2 Output overload

1.3 Connections

SIM983 Scaling Amplifier
An overload condition is recognized and the input OVLD LED is
activated if the absolute value of the voltage applied to the input
exceeds certain limits. These limits are typically
between
10 4 V
V
min
The overloaded state is also recognized, and the input overload LED
activated, if the sum of the input voltage and the commanded o set,
V
V
exceeds these limits. To distinguish between the two input
in
ofs
overload possibilities, use the command OVLD? . The overload LED
stays on for a minimum of 50 ms; after this time it turns o if the
overload condition has ceased.
An overload condition is recognized and the output OVLD LED is
activated if the absolute value G
Section 1.2.4.1. The overload LED stays on for a minimum of 50 ms;
after this time it turns o if the overload condition has ceased.
For a discussion of the front and rear BNC connections, see Sec-
tion 2.1. The SIM interface connector is discussed in Section 1.6.1.
9 9 V
9 9 V
V
max
V
(V
) exceeds the limits in
in
ofs
1 – 5