Dispenser calibration Dispenser pump 1
V:
100 ml
t:
-
OK
+
Enter the amount measured in field "V:".
The throughput P is calculated in ml/min. by the controls.
Press the "Save" button.
The measured value is assigned to the pump.
Reattach the hose.
Dispensing throughput (D1 to D13)
The throughput of the connected dispensing pumps can be set from
0 ml/min to 9999 ml/min in 1 ml/min steps.
Access the following content:
Supervisor level
Settings
Dispensing throughput
Dispensing correction factor
OK
Select "Dispensing throughput".
The following will appear on the display:
Settings
Dispensing throughput
Dispenser pump 1
: 100,0 ml/min
Dispenser pump 2
:
100,0 ml/min
OK
Options:
– No
– 1 to 9999.0 ml/min
30 s
P: 200,0 ml/min
0/1
Settings option
99